Skip to main content
JDRS · Open access journal

Original Article · Open access

Biomechanical comparison of three fixation methods under axial loading in proximal humerus fractures with medial metaphyseal defect

PDF Cite this article

Muhlik Akyürek

Department of Orthopedics and Traumatology, Maria-josef Hospital, Greven, Germany

ORCID: 0000-0002-4907-8840

Emre Koraman

Department of Orthopedics and Traumatology, Pendik Bölge Hospital, İstanbul, Türkiye

ORCID: 0000-0003-2715-2875

Aykut Çelik

Department of Orthopedics and Traumatology, Beykoz State Hospital, İstanbul, Türkiye

ORCID: 0000-0002-5910-0829

Yusuf Iyetin

Department of Orthopedics and Traumatology, Pendik Bölge Hospital, İstanbul, Türkiye

ORCID: 0000-0001-5530-7729

Clemens Kösters

Department of Orthopedics and Traumatology, Maria-josef Hospital, Greven, Germany

ORCID: 0000-0002-9150-7282

Ismail Türkmen

Department of Orthopedics and Traumatology, Klinikum Osnabrück GmbH Center of Musculoskeletal Surgery, Osnabrück, Germany

ORCID: 0000-0001-7594-1989

Abstract

Objectives: This study aims to compare the biomechanical performance of three different surgical fixation techniques in the treatment of proximal humerus fractures with medial metaphyseal defects.

Materials and methods: A total of 24 synthetic humerus bone models were used and equally divided into three groups (n = 8 per group). A standardized unstable proximal humerus fracture model with a medial metaphyseal cortical defect extending to the surgical neck was created using a three-dimensional (3D)-printed osteotomy guide. Group 1 received fixation with a lateral anatomical locking plate. Group 2 was treated with a combination of a lateral anatomical plate and a medial buttress plate (dual plating). Group 3 underwent fixation with an intramedullary nail (IMN); in this group, four specimens had distal locking with an endopin (Group 3a), and the other four with static screws (Group 3b). All specimens were subjected to axial loading until failure. Forces at the onset of failure and complete failure were recorded, and fracture patterns were documented.

Results: In Groups 1 and 2, transverse fractures consistently occurred at the level of the most distal screw of the lateral plate. In Group 3, failure was observed either proximally or distally at the nail tip, including butterfly fragment formation and metaphyseal collapse. Group 2 exhibited the highest resistance to axial loading, followed by Group 1 and Group 3, with statistically significant differences between all groups (p = 0.043, p = 0.0003, p < 0.00001). No significant difference was found between subgroups 3a and 3b (p > 0.05).

Conclusion: Our study results indicate that double plating provides the greatest axial stability in proximal humerus fractures with medial metaphyseal defects, supporting its use in fracture patterns with medial column deficiency. However, as fixation choice should be guided by patient-specific factors and surgical feasibility, these findings should be interpreted in the context of experimental conditions and loading limitations.

Declarations

Author Contributions

M.A.: Idea/concept, design, writing the article, other (performed surgical procedures); E.K.: Data collection and processing, analysis, writing the article; A.Ç.: Other (performed surgical procedures), data collection, references; Y.I.: Design, literature review; C.K.: Critical review, control/supervision; I.T.: Materials, idea/ concept, control/supervision.

Conflict of Interest

The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.

Financial Disclosure

The authors received no financial support for the research and/or authorship of this article.

Data Sharing Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

AI Disclosure

The authors declare that artificial intelligence (AI) tools were not used, or were used solely for language editing, and had no role in data analysis, interpretation, or the formulation of conclusions. All scientific content, data interpretation, and conclusions are the sole responsibility of the authors. The authors further confirm that AI tools were not used to generate, fabricate, or ‘hallucinate’ references, and that all references have been carefully verified for accuracy.

References

  1. Iglesias-Rodríguez S, Domínguez-Prado DM, García-Reza A, Fernández-Fernández D, Pérez-Alfonso E, García-Piñeiro J, et al. Epidemiology of proximal humerus fractures. J Orthop Surg Res 2021;16:402. DOI: 10.1186/s13018-021-02551-x.
  2. Shi Y, Zhou J. Comparative finite element analysis: internal fixation plate versus intramedullary nail for neer type III-VI proximal humeral fractur. J Orthop Surg Res 2025;20:4. DOI: 10.1186/s13018-024-05418-z.
  3. Song LM, Wang GX, Wang L. Comparison between CFRPEEK and titanium plate for proximal humeral fracture: A meta-analysis. Jt Dis Relat Surg 2024;35:483-90. DOI: 10.52312/jdrs.2024.1611.
  4. Park MC, Murthi AM, Roth NS, Blaine TA, Levine WN, Bigliani LU. Two-part and three-part fractures of the proximal humerus treated with suture fixation. J Orthop Trauma 2003;17:319-25. DOI: 10.1097/00005131-200305000-00001.
  5. Robinson CM, Page RS, Hill RM, Sanders DL, CourtBrown CM, Wakefield AE. Primary hemiarthroplasty for treatment of proximal humeral fractures. J Bone Joint Surg Am 2003;85:1215-23. DOI: 10.2106/00004623-200307000-00006.
  6. Strømsøe K. Fracture fixation problems in osteoporosis. Injury 2004;35:107-13. DOI: 10.1016/j.injury.2003.08.019.
  7. Korsić M, Grazio S. Non-hip peripheral osteoporotic fractures: Epidemiology and significance. Arh Hig Rada Toksikol 2008;59:53-8. DOI: 10.2478/10004-1254-59-2008-1850.
  8. Gardner MJ, Boraiah S, Helfet DL, Lorich DG. The anterolateral acromial approach for fractures of the proximal humerus. J Orthop Trauma 2008;22:132-7. DOI: 10.1097/BOT.0b013e3181589f8c.
  9. Maier D, Jaeger M, Izadpanah K, Strohm PC, Suedkamp NP. Proximal humeral fracture treatment in adults. J Bone Joint Surg Am 2014;96:251-61. DOI: 10.2106/JBJS.L.01293.
  10. Park SG, Ko YJ. Medial buttress plating for humerus fractures with unstable medial column. J Orthop Trauma 2019;33:e352-9. DOI: 10.1097/BOT.0000000000001515.
  11. Wang F, Wang Y, Dong J, He Y, Li L, Liu F, et al. A novel surgical approach and technique and short-term clinical efficacy for the treatment of proximal humerus fractures with the combined use of medial anatomical locking plate fixation and minimally invasive lateral locking plate fixation. J Orthop Surg Res 2021;16:29. DOI: 10.1186/s13018-020-02094-7.
  12. Bastian JD, Hertel R. Initial post-fracture humeral head ischemia does not predict development of necrosis. J Shoulder Elbow Surg 2008;17:2-8. DOI: 10.1016/j.jse.2007.03.026.
  13. Edwards SL, Wilson NA, Zhang LQ, Flores S, Merk BR. Two-part surgical neck fractures of the proximal part of the humerus. A biomechanical evaluation of two fixation techniques. J Bone Joint Surg Am 2006;88:2258-64. DOI: 10.2106/JBJS.E.00757.
  14. Huff LR, Taylor PA, Jani J, Owen JR, Wayne JS, Boardman ND 3rd. Proximal humeral fracture fixation: A biomechanical comparison of two constructs. J Shoulder Elbow Surg 2013;22:129-36. DOI: 10.1016/j.jse.2012.01.003.
  15. He Y, Zhang Y, Wang Y, Zhou D, Wang F. Biomechanical evaluation of a novel dualplate fixation method for proximal humeral fractures without medial support. J Orthop Surg Res 2017;12:72. DOI: 10.1186/s13018-017-0573-4.
  16. He Y, He J, Wang F, Zhou D, Wang Y, Wang B, et al. Application of additional medial plate in treatment of proximal humeral fractures with unstable medial column: A finite element study and clinical practice. Medicine (Baltimore) 2015;94:e1775. DOI: 10.1097/MD.0000000000001775.
  17. Boadi PJ, Da Silva A, Mizels J, Joyce CD, Anakwenze OA, Klifto CS, et al. Intramedullary versus locking plate fixation for proximal humerus fractures: Indications and technical considerations. JSES Rev Rep Tech 2024;4:615-24. DOI: 10.1016/j.xrrt.2024.01.001.
  18. Jang Y, Kim D. Biomechanical study of Proximal humeral fracture fixation: Locking plate with medial support screw vs. locking plate with intramedullary fibular graft. Clin Biomech (Bristol) 2021;90:105510. DOI: 10.1016/j.clinbiomech.2021.105510.
  19. Jabran A, Peach C, Ren L. Biomechanical analysis of plate systems for proximal humerus fractures: A systematic literature review. Biomed Eng Online 2018;17:47. DOI: 10.1186/s12938-018-0479-3.
  20. Padegimas EM, Zmistowski B, Lawrence C, Palmquist A, Nicholson TA, Namdari S. Defining optimal calcar screw positioning in proximal humerus fracture fixation. J Shoulder Elbow Surg 2017;26:1931-7. DOI: 10.1016/j.jse.2017.05.003.
  21. Lim JH, Hwang J, Kim S, Kim MS. Clinical and radiographic results of locking plate with medial support screw in Proximal Humerus fracture - the more, the better? BMC Musculoskelet Disord 2024;25:580. DOI: 10.1186/s12891-024-07700-x.
  22. Hettrich CM, Boraiah S, Dyke JP, Neviaser A, Helfet DL, Lorich DG. Quantitative assessment of the vascularity of the proximal part of the humerus. J Bone Joint Surg Am 2010;92:943-8. DOI: 10.2106/JBJS.H.01144.
  23. Hertel R, Hempfing A, Stiehler M, Leunig M. Predictors of humeral head ischemia after intracapsular fracture of the proximal humerus. J Shoulder Elbow Surg 2004;13:427-33. DOI: 10.1016/j.jse.2004.01.034.
  24. Zhang Y, Wan L, Zhang L, Yan C, Wang G. Reduction and fixation of proximal humeral fracture with severe medial instability using a small locking plate. BMC Surg 2021;21:387. DOI: 10.1186/s12893-021-01388-9.
  25. Barlow JD, Logli AL, Steinmann SP, Sems SA, Cross WW, Yuan BJ, et al. Locking plate fixation of proximal humerus fractures in patients older than 60 years continues to be associated with a high complication rate. J Shoulder Elbow Surg 2020;29:1689-94. DOI: 10.1016/j.jse.2019.11.026.
  26. Theopold J, Marquaß B, Fakler J, Steinke H, Josten C, Hepp P. The bicipital groove as a landmark for reconstruction of complex proximal humeral fractures with hybrid double plate osteosynthesis. BMC Surg. 2016;16:10.
  27. Lee CW, Shin SJ. Prognostic factors for unstable proximal humeral fractures treated with locking-plate fixation. J Shoulder Elbow Surg 2009;18:83-8. DOI: 10.1016/j.jse.2008.06.014.
  28. Owsley KC, Gorczyca JT. Fracture displacement and screw cutout after open reduction and locked plate fixation of proximal humeral fractures [corrected]. J Bone Joint Surg Am 2008;90:233-40. DOI: 10.2106/JBJS.F.01351.
  29. Gardner MJ, Weil Y, Barker JU, Kelly BT, Helfet DL, Lorich DG. The importance of medial support in locked plating of proximal humerus fractures. J Orthop Trauma 2007;21:185- 91. DOI: 10.1097/BOT.0b013e3180333094.
  30. Muccioli C, Chelli M, Caudal A, Andreani O, Elhor H, Gauci MO, et al. Rotator cuff integrity and shoulder function after intra-medullary humerus nailing. Orthop Traumatol Surg Res 2020;106:17-23. DOI: 10.1016/j.otsr.2019.11.004.
  31. Tasci M, Turkmen İ, Celik H, Akcal MA, Şekerci R, Keles N, et al. InSafeLock humeral nail provides a safe application for proximal and distal locking screws with distal endopin - An anatomical study. Orthop Traumatol Surg Res 2019;105:1005-11. DOI: 10.1016/j.otsr.2019.04.014.
  32. Michel PA, Katthagen JC, Heilmann LF, Dyrna F, Schliemann B, Raschke MJ. Biomechanics of upper extremity double plating. Z Orthop Unfall 2020;158:238-44. DOI: 10.1055/a0862-6334.
  33. Dey Hazra RO, Blach RM, Ellwein A, Katthagen JC, Lill H, Jensen G. Latest trends in the current treatment of proximal humeral fractures - an analysis of 1162 cases at a level-1 trauma centre with a special focus on shoulder surgery. Z Orthop Unfall 2022;160:287-98. DOI: 10.1055/a-1333-3951.