Skip to main content
JDRS · Open access journal

Case Report · Open access

Wire-stem fretting and galvanic corrosion as an underrecognized cause of metallosis after hip arthroplasty: A rare case report

PDF Cite this article

Wei-Cheng Chiu

Department of Orthopedics, Cardinal Tien Hospital, New Taipei City, Taiwan Department of Surgery, Cardinal Tien Hospital, New Taipei City, Taiwan School of Medicine, Fu-Jen Catholic University, New Taipei City, Taiwan

ORCID: 0009-0005-0694-7064

Hui-Jung Yu

School of Medicine, Fu-Jen Catholic University, New Taipei City, Taiwan Department of Pathology, Cardinal Tien Hospital, New Taipei City, Taiwan

ORCID: 0009-0006-4629-9732

Wei-Zhi Liao

Department of Orthopedics, Cardinal Tien Hospital, New Taipei City, Taiwan

ORCID: 0009-0008-0560-9911

Abstract

Although metallosis after hip arthroplasty is most often attributed to bearing wear or trunnion corrosion, auxiliary fixation devices may also serve as a potential source. In this article, we report a rare case in which direct contact between a stainless-steel cerclage wire and a titanium femoral stem produced localized metallic debris and a peri-stem pseudotumor, consistent with fretting and galvanic corrosion. Ten years after index total hip arthroplasty complicated by an intraoperative femoral fracture treated with cerclage, the patient presented with groin pain and a cystic lesion with solid components adjacent to the femoral stem. Revision surgery revealed black-stained tissue concentrated at the wire-stem interface. Histological examination confirmed metallic debris with chronic inflammation. After extensive debridement and revision to a distally fixed stem, pain on the Visual Analog Scale (VAS) decreased from 5 to 1, and the Harris Hip Score (HHS) improved from 70 to 93 at six months. In conclusion, this case serves as a reminder that even auxiliary fixation devices, such as stainless-steel cerclage wires placed for intraoperative fracture stabilization, can generate metallic debris and lead to metallosis when they come into contact with titanium stems. Surgeons, therefore, should avoid direct contact between stainless-steel and titanium components, or otherwise interpose an insulating layer when cerclage fixation is required.

Declarations

Author Contributions

W.C.C.: Conceptualization, writing-original draft; W.C.C., H.J.Y.: Methodology; W.C.C., H.J.Y., W.Z.L.: Investigation, resources; W.C.C., W.Z.L.: Writing-review & editing; W.Z.L.: Supervision. All authors had full access to the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis.

Conflict of Interest

The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.

Financial Disclosure

The authors received no financial support for the research and/or authorship of this article.

Data Sharing Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

AI Disclosure

The authors declare that artificial intelligence (AI) tools were not used, or were used solely for language editing, and had no role in data analysis, interpretation, or the formulation of conclusions. All scientific content, data interpretation, and conclusions are the sole responsibility of the authors. The authors further confirm that AI tools were not used to generate, fabricate, or ‘hallucinate’ references, and that all references have been carefully verified for accuracy.

References

  1. Ude CC, Esdaille CJ, Ogueri KS, Ho-Man K, Laurencin SJ, Nair LS, et al. The mechanism of metallosis after total hip arthroplasty. Regen Eng Transl Med 2021;7:247-61. DOI: 10.1007/s40883-021-00222-1.
  2. Stoops TK, Layuno-Matos JG, Simon P, Gustke KA, Bernasek TL. Metal-on-metal total hip arthroplasties: Why do they fail? Arthroplast Today 2025;33:101662. DOI: 10.1016/j.artd.2025.101662.
  3. Orfanos G, Wainwright TW, Middleton RG. An economic analysis of metal-on-metal versus ceramic-on-ceramic hip resurfacing: Costs and clinical implications. Prosthesis 2024;6:1586-90. DOI: 10.3390/prosthesis6060114.
  4. Sangani K, Ramesh G, Chakrapani A. Metallosis following non-metal-on-metal hip arthroplasty: A case report and review. J Orthop Case Rep 2024;14:111-5. DOI: 10.13107/jocr.2024.v14.i09.4748.
  5. Stefanini N, Morandi Guaitoli M, Pilla F, Faldini C. Managing total hip arthroplasty failure due to ceramic liner fracture and cup perforation: Revision with bone grafting through direct anterior approach. BMJ Case Rep 2025;18:e263855. DOI: 10.1136/bcr-2024-263855.
  6. Skjöldebrand C, Tipper JL, Hatto P, Bryant M, Hall RM, Persson C. Current status and future potential of wearresistant coatings and articulating surfaces for hip and knee implants. Mater Today Bio 2022;15:100270. DOI: 10.1016/j.mtbio.2022.100270.
  7. Schoon J, Hesse B, Rakow A, Ort MJ, Lagrange A, Jacobi D, et al. Metal-specific biomaterial accumulation in human peri-implant bone and bone marrow. Adv Sci (Weinh) 2020;7:2000412. DOI: 10.1002/advs.202000412.
  8. Tandon M, Chetla N, Hodges J, Koul A, Dharia S, Shah D, et al. Mechanical considerations and clinical implications of joint arthroplasty metallosis. Cureus 2024;16:e76592. DOI: 10.7759/cureus.76592.
  9. Günther KP, Deckert S, Lützner C, Lange T, Schmitt J, Postler A, et al. Total hip replacement for osteoarthritisevidence-based and patient-oriented indications. Dtsch Arztebl Int 2021;118:730-6. DOI: 10.3238/arztebl.m2021.0323.
  10. Urish KL, Giori NJ, Lemons JE, Mihalko WM, Hallab N. Trunnion corrosion in total hip arthroplasty-basic concepts. Orthop Clin North Am 2019;50:281-8. DOI: 10.1016/j.ocl.2019.02.001.
  11. Zhang L, Haddouti EM, Welle K, Burger C, Kabir K, Schildberg FA. Local cellular responses to metallic and ceramic nanoparticles from orthopedic joint arthroplasty implants. Int J Nanomedicine 2020;15:6705-20. DOI: 10.2147/IJN.S248848.
  12. Radzik B, Bijukumar D, Cheng KY, Badhe RV, Barba M, Mathew MT. The role of fretting-frequency on the damage modes of THR modular junction: In-vitro study. Mater Sci Eng C Mater Biol Appl 2021;126:112128. DOI: 10.1016/j.msec.2021.112128.
  13. Migliorini F, Pilone M, Bell A, Merfort R, Giorgino R, Maffulli N. Serum cobalt and chromium concentration following total hip arthroplasty: A Bayesian network meta-analysis. Sci Rep 2023;13:6918. DOI: 10.1038/s41598-023-34177-w.
  14. Venkatraman V, Wong MK, Shalita C, Parente B, Lad SP. Cobalt-induced toxicity and spasticity secondary to hip arthroplasty: Case report and review of the literature. Cureus 2020;12:e12368. DOI: 10.7759/cureus.12368.
  15. Umar M, Jahangir N, Faisal Khan M, Saeed Z, Sultan F, Sultan A. Cobalt cardiomyopathy in hip arthroplasty. Arthroplast Today 2019;5:371-5. DOI: 10.1016/j.artd.2019.04.010.
  16. Mohandas P, Magadum SD, Prasad JR, Karthick M, Singh UK, Purushothaman B. Unexplained pseudotumor formation in ceramic-on-ceramic hip bearings: A case report. J Orthop Case Rep 2025;15:18-23. DOI: 10.13107/jocr.2025.v15.i08.5868.
  17. Wang X, Li Y, Feng Y, Cheng H, Li D. Macrophage polarization in aseptic bone resorption around dental implants induced by Ti particles in a murine model. J Periodontal Res 2019;54:329-38. DOI: 10.1111/jre.12633.
  18. Terhune EB, Sutter EG, Balkissoon R, Pallante G, Specht LM, Leikin J, et al. Cobalt toxicity from third-body wear following revision of fractured ceramic-on-ceramic hip bearings with a metal articulation: The fractured ceramic syndrome. Bone Joint J 2025;107-B:92-100. DOI: 10.1302/0301-620X.107B6.BJJ-2024-1114.R1.
  19. Blackmon J, Blackmon L, Goode C, Douthit N. Systemic cobalt toxicity secondary to metal-on-metal prosthetic hip replacement: A case report. J Gen Intern Med 2024;39:133-7. DOI: 10.1007/s11606-023-08490-z.
  20. Ahmed A, Grabie Y, Loeffler J, Buchen Y, Acharya S. Reactive peri-arthroplasty pseudotumors: A rare cause of hip pain and iatrogenic sciatica. Cureus 2023;15:e47322. DOI: 10.7759/cureus.47322.
  21. Serhan H, Slivka M, Albert T, Kwak SD. Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern? Spine J 2004;4:379-87. DOI: 10.1016/j.spinee.2003.12.004.
  22. Høl PJ, Mølster A, Gjerdet NR. Should the galvanic combination of titanium and stainless steel surgical implants be avoided? Injury 2008;39:161-9. DOI: 10.1016/j.injury.2007.07.015.
  23. Dudko S, Kusz D, Kopeć K, Wojciechowski P, Kwiatkowska K, Niemiec D. Pseudotumor as a complication of total hip replacement. Ortop Traumatol Rehabil 2022;24:273-80. DOI: 10.5604/01.3001.0015.9991.
  24. Brun OC, Maansson L. Fractures of the greater trochanter following total hip replacement. Hip Int 2013;23:143-6. DOI: 10.5301/HIP.2013.10813.
  25. Inoue D, Kabata T, Kajino Y, Ohmori T, Yamamuro Y, Tsuchiya H. Risk factor analysis on perioperative greater trochanteric fracture of total hip arthroplasty via anterolateral approach. Arch Orthop Trauma Surg 2023;143:4519-25. DOI: 10.1007/s00402-022-04715-8.
  26. Kim HS, Lee DK, Mun KU, Moon DH, Kim CH. What is the best treatment choice for concomitant ipsilateral femoral neck and intertrochanteric fracture? A retrospective comparative analysis of 115 consecutive patients. J Pers Med 2022;12:1908. DOI: 10.3390/jpm12111908.
  27. Sanderson B, Washburn F, Allison D. Triple surgical fixation technique for an isolated greater trochanter fracture in an amateur weightlifter. J Am Acad Orthop Surg Glob Res Rev 2022;6:e21.00127. DOI: 10.5435/JAAOSGlobal-D-21-00127.
  28. Donegan D, Huo M, Leslie M. Tension band wiring for Trochanteric femoral fracture. AO Surgery Reference. 2022, [cited, 25.11.2025]. Available from: https://surgeryreference. aofoundation.org/orthopedic-trauma/periprostheticfractures/hip/trochanteric-femoral-fracture/tension-bandwiring.